    內窺鏡設備維修的專業性展現:從技術規范到服務閉環的全流程解析

    來源:http://www.xiuchezu.com/ 發布時間:2025-06-09 瀏覽量:0

      內窺鏡作為醫療診斷與工業檢測的核心設備,其維修服務的技術門檻與規范性要求極高。從故障診斷到性能恢復,每個環節都需遵循嚴謹的技術標準與操作規范,才能確保設備安全與檢測精度。以下從技術、流程、管理三個維度,解析內窺鏡設備維修專業性的具體體現。

      As the core equipment for medical diagnosis and industrial testing, endoscopes have extremely high technical barriers and regulatory requirements for maintenance services. From fault diagnosis to performance recovery, every step must follow rigorous technical standards and operating procedures to ensure equipment safety and detection accuracy. Below, we will analyze the specific manifestations of the professionalism of endoscopic equipment maintenance from three dimensions: technology, process, and management.

      一、技術能力:精準診斷與規范修復

      1、 Technical Capability: Accurate Diagnosis and Standardized Repair

      故障定位技術

      Fault location technology

      維修人員需掌握光學系統、機械結構、電子電路的綜合檢測能力。例如,針對圖像模糊問題,需通過光通量測試儀檢測導光束衰減率,若超過30%則需更換光纖組件;對彎曲部卡滯故障,需使用內窺鏡專用測力計測量操作角阻力,標準值應控制在1.5N以內。

      Maintenance personnel need to master the comprehensive testing ability of optical systems, mechanical structures, and electronic circuits. For example, to address the issue of image blur, it is necessary to use a luminous flux tester to detect the attenuation rate of the guide beam. If it exceeds 30%, the fiber optic components need to be replaced; For the stuck fault of the bent part, a specialized endoscope force gauge should be used to measure the operating angle resistance, and the standard value should be controlled within 1.5N.

      部件修復工藝

      Component repair process

      光學鏡頭修復:采用激光干涉儀檢測透鏡面型精度,面型偏差PV值超過λ/4(λ=632.8nm)時需進行拋光處理,使用氧化鈰拋光粉配合聚氨酯拋光墊,修復后需通過MTF(調制傳遞函數)測試。

      Optical lens repair: Laser interferometer is used to detect the surface accuracy of the lens. When the surface deviation PV value exceeds λ/4 (λ=632.8nm), polishing treatment is required. Cerium oxide polishing powder is used in combination with polyurethane polishing pad. After repair, MTF (modulation transfer function) test is required.

      蛇骨組件維護:對彎曲橡膠進行邵氏硬度檢測,硬度低于A60時需整體更換,裝配時采用分段注膠工藝,確保彎曲半徑≤50mm時無褶皺。

      Snake bone component maintenance: Perform Shore hardness testing on the bent rubber. If the hardness is lower than A60, it needs to be replaced as a whole. During assembly, use segmented injection molding technology to ensure that there are no wrinkles when the bending radius is ≤ 50mm.

      電路板維修:使用熱風槍進行BGA芯片焊接,溫度曲線需精準控制:預熱段120℃/60秒,回流段245℃/40秒,冷卻段30℃/秒,配合X-Ray檢測焊點空洞率≤5%。

      Circuit board repair: Use a hot air gun for BGA chip soldering, and the temperature curve needs to be accurately controlled: preheating section at 120 ℃/60 seconds, reflow section at 245 ℃/40 seconds, cooling section at 30 ℃/second, combined with X-ray inspection to detect solder joint void rate ≤ 5%.

    20200408042119178

      密封性驗證

      Sealing verification

      對醫用內窺鏡,需進行氣密性檢測,壓力值設定為30kPa,保壓時間10分鐘,泄漏率≤0.05mL/min。工業內窺鏡則需進行水壓試驗,壓力為額定工作壓力的1.5倍,持續時間30分鐘,無滲漏現象。

      For medical endoscopes, airtightness testing is required, with a pressure value set at 30kPa, a holding time of 10 minutes, and a leakage rate of ≤ 0.05mL/min. Industrial endoscopes require a water pressure test, with a pressure of 1.5 times the rated working pressure and a duration of 30 minutes, without any leakage.

      二、流程管理:標準化作業與質量追溯

      2、 Process Management: Standardized Operations and Quality Traceability

      五級驗收制度

      Five level acceptance system

      預檢:通過設備自檢程序獲取故障代碼,結合使用記錄分析故障頻譜。

      Pre check: Obtain fault codes through the device self-test program, and analyze the fault spectrum in combination with usage records.

      拆解:在潔凈度等級ISO 5級的環境下進行,關鍵部件使用防靜電鑷子操作。

      Disassembly: Conducted in an environment with a cleanliness level of ISO 5, key components are operated using anti-static tweezers.

      修復:執行《內窺鏡維修技術規范》,關鍵工序雙人復核。

      Repair: Implement the "Technical Specification for Endoscope Maintenance" and double check key processes.

      組裝:按逆向工序回裝,扭矩值采用數顯扭矩扳手控制,誤差±2%。

      Assembly: Reinstall according to the reverse process, using a digital torque wrench to control the torque value with an error of ± 2%.

      終檢:模擬實際工況進行全功能測試,測試點覆蓋所有操作模式。

      Final inspection: Simulate actual working conditions for full functional testing, with testing points covering all operating modes.

      溯源管理系統

      Traceability Management System

      建立設備“一機一檔”數據庫,記錄序列號、維修歷史、更換部件批次等信息。關鍵部件實施二維碼追溯,掃描可獲取供應商資質、質檢報告、安裝日期等數據。

      Establish a "one machine, one file" database for equipment, recording information such as serial numbers, maintenance history, and replacement part batches. Implement QR code traceability for key components, and scan to obtain supplier qualifications, quality inspection reports, installation dates, and other data.

      預防性維護方案

      Preventive maintenance plan

      根據設備使用頻次制定維護周期:高頻使用設備(日檢測量>50次)每3個月進行光學系統校準,每6個月更換密封圈;低頻設備每年進行全面性能檢測。

      Develop maintenance cycle based on equipment usage frequency: For high-frequency equipment (with daily inspection and measurement>50 times), optical system calibration should be performed every 3 months, and sealing rings should be replaced every 6 months; Low frequency equipment undergoes comprehensive performance testing annually.

      本文由內窺鏡設備維修友情奉獻.更多有關的知識請點擊:http://www.xiuchezu.com我們將會對您提出的疑問進行詳細的解答,歡迎您登錄網站留言.

      This article is a friendly contribution from the maintenance of endoscopic equipment For more information, please click: http://www.xiuchezu.com We will provide detailed answers to your questions. You are welcome to log in to our website and leave a message

    s
    新聞資訊
    聯系我們
    • 公司公司:匠仁醫療設備有限公司
    • 匠仁醫療山東濟南維修部
    • 聯系人:樊經理
    • 電話:13153199508
    • 地址:山東省濟南市槐蔭區美里東路3000號德邁國際信息產業園6號樓101-2室
    • 匠仁醫療湖南長沙維修部
    • 聯系人:李經理
    • 電話:13873135765
    • 地址:湖南省長沙市雨花區勞動東路820號恒大綠洲14棟2409室
    新聞資訊

    樊經理:13153199508     李經理:13873135765

    公司地址:山東省濟南市槐蔭區美里東路3000號德邁國際信息產業園6號樓101-2室  湖南省長沙市雨花區勞動東路820號恒大綠洲14棟2409室

         

    備案號:魯ICP備2023027194號-1    魯公網安備 37010402441281號